How to Read an HPLC Chromatogram: A Practical Guide
Share
Open the analytical report for a research peptide and there's usually a graph in there: a flat line, one tall spike and maybe a few little bumps. Once you know how to read an HPLC chromatogram, that graph stops being decoration and starts being evidence. It's where the headline purity figure comes from, and it often tells you more than the figure does on its own.

This one's for researchers who want to understand the result, not run the instrument.
What you're actually looking at
HPLC (high-performance liquid chromatography) pushes a dissolved sample through a packed column. Different components travel through at different speeds depending on how strongly they cling to the packing and the solvent. As each one comes out the other end, a detector records a signal. For peptides that's usually UV at around 214 to 220 nm, where the peptide bond absorbs. The chromatogram is that signal plotted over time.
The horizontal axis is time, normally in minutes from when the sample went in. The point where a peak appears is its retention time, and under the same conditions the same compound should turn up at a consistent time. Change the column, solvent or gradient, though, and the times shift, so comparing retention times between different methods doesn't tell you much. The vertical axis is the detector's response, often in milli-absorbance units (mAU). Taller means a stronger signal at that moment.

How to read an HPLC chromatogram's peaks and purity figure
Every peak is something that came off the column. In a clean sample you'd expect one dominant main peak and maybe a few much smaller ones from related impurities, like truncated or modified sequences. The software measures the area under each peak, and it's area, not height, that purity calculations use. A short, broad peak and a tall, skinny one can represent similar amounts. That table under the graph with retention times and “area %” is the software's integration of each peak.
The usual purity figure is area percent: the main peak's area divided by the total area of all the integrated peaks, times 100. Say the main peak comes in at 980 units and three small peaks add up to 20. That's 98%. (Made-up numbers, just to show the sum.) It's a useful figure, but it only counts what the detector saw at that wavelength and what the software integrated. Anything that doesn't absorb there, or never made it off the column, is invisible. Our piece on what 98% and 99% purity mean goes further into that.

The baseline and the shoulders
The flat line between peaks is the baseline. When it's steady and quiet, small peaks are easy to see and measure. When it's noisy or drifting, it's hard to tell a real impurity from background, and small peaks can end up under- or over-counted.
Then look closely at the main peak itself. Resolution is how cleanly two neighbouring peaks are separated. If an impurity comes off very close to the main peak, the two can blur into one peak with a shoulder on it, and the software may count the impurity as part of the main peak. That inflates the purity figure. A shoulder or a split top is worth a second look.
Well-run labs also check the instrument before or during a run using reference standards or test mixtures. These system suitability checks confirm that things like retention time consistency, peak shape and resolution are within limits on the day. Plenty of summary certificates don't mention them, but when they're there, they add confidence the run itself was behaving. Our guide to analytical method validation explains how that differs from validating a method.
98% doesn't mean 98% of the vial
This is the big one. “98% purity” describes the relative proportion of the components the detector saw, not the absolute amount of material in the vial. Freeze-dried peptides also hold water and counter-ions (acetate or trifluoroacetate, for example) that don't show up as peaks in a UV chromatogram. How much peptide is actually in there is a separate measurement, usually reported as assay or content.
When we read one ourselves, we go roughly in this order. Check the product, reference and test date match the material. Note the method details if they're given. Find the main peak and its retention time. Have a look at the baseline, then check for shoulders near the main peak. Read the peak table and make sure the area % matches the purity figure quoted. Then pair it with the identity result, because purity without identity is only half the story. Our peptide mass spectrometry guide covers that side.
For a number of our products we publish third-party COAs with verification links on our COAs page. If the one you want isn't up, flick us an email at support@ascendaus.com and we'll let you know what's available.
Frequently asked questions
What does retention time mean on an HPLC chromatogram?
It's how long a component takes to get through the column to the detector. Under the same method conditions, the same compound should show up at a consistent time.
How is HPLC purity calculated?
Most reports use area percent: the main peak's area divided by the total area of all integrated peaks, times 100.
Why is peak area used instead of peak height?
Area captures a component's whole signal, while height depends on the peak's shape. A short, broad peak and a tall, narrow one can represent similar amounts.
Does 98% HPLC purity mean 98% of the vial is peptide?
No. Area percent compares the components the detector saw. Water and counter-ions don't appear as UV peaks, and the actual amount of peptide is a separate assay or content measurement.
What does a shoulder on the main peak mean?
Usually that another component is coming off very close to the main peak and isn't fully separated. It can make the purity figure look better than it is.